A test rig is the hardware that lets you answer a test requirement. It presents the device, or part of it, the way real use would, applies the load, motion, pressure or flow the method calls for, and measures the response in a way you can stand behind later. Sometimes that’s a fixture in a universal force tester. Quite often it’s more: a rig with its own actuation, a surrogate for the tissue or mating product the device meets in use, or an instrumented version of the device itself. We design and build these in our own workshop near Bristol, so this post covers how test rig design works in practice, written from the bench. The case for bench testing in the first place is in our post on custom test rigs and bench testing.
Test rig design starts with the requirement
Every rig we build starts with a sentence from somewhere: a clause in a standard, a line in the design inputs, or a claim in the instructions for use. “The cap removal force shall not exceed X newtons”, or “the joint shall withstand Y cycles without loss of function”. Most rig problems trace back to skipping this step, because a rig built before the requirement is pinned down tends to measure something adjacent to the question.
So the first thing we write is a short test requirement statement: what is being measured, in what units, under what conditions, with what pass/fail criterion, and which document the result will feed (usually a design verification report, which then sits in the technical file). Everything in the rig follows from that statement.
What a test rig can be
Sometimes a rig is a clamp in a universal tester and nothing more. Most requirements worth testing need something else, and rigs tend to take one of these shapes.
The rig provides what the device acts on. A device that pushes against tissue, grips a limb or mates with another product can’t be tested in free air, so the rig supplies a surrogate. That might be a silicone skin analogue at a representative hardness, a foam or gel block standing in for soft tissue, an anatomical form, or a sample of the actual mating part. The surrogate’s stiffness and geometry shape the result, so we choose it against the requirement and document it in the method like any other piece of apparatus.
The rig is part of the device. Early in development you often need to test a subassembly before the rest of the device exists: a drive spring, a latch, a hinge, a seal. The rig stands in for the missing parts and reproduces the constraints and reactions the subassembly would see in the full product. Done well, this verifies the risky mechanism months before a full build. Done carelessly, the rig constrains the part differently to the device and the result doesn’t transfer.
The rig has its own actuation and sensing. A fluid-handling device needs a circuit: a pump or pressure source, representative tubing runs, pressure and flow measurement, and somewhere for the fluid to go. A cycling test often needs its own pneumatic or motor-driven actuator, because it would otherwise tie up a universal tester for days. An instrumented build of the device, with a sensor buried where you could never see in use, is a rig in its own right, and sometimes it’s the only way to measure what the requirement asks about.
In each case the rig’s job is to reproduce the conditions the requirement describes, and only those. Whatever the rig replaces (tissue, mating parts, the rest of the device, the user’s hand) has to be justified, because it’s the first thing a reviewer will probe.
Holding the sample
Holding the sample sounds trivial and almost never is. The rig has to hold the device or subassembly firmly enough that nothing slips, without deforming it in a way that changes the result, and it has to present every sample identically, because repeatability between samples is what makes the data usable. For a subassembly that also means supporting it where the rest of the device would, and nowhere else.
We make the holding parts to match the device geometry: a printed nest that cradles a moulded housing, a turned collet that grips a cylindrical body, a clamp with a silicone liner where the part would otherwise mark. First iterations are usually printed because they’re quick to revise, and we move to metal where stiffness matters or the part will see thousands of cycles. The check is always the same. If we hand the rig and ten samples to someone else, do they load each one identically without asking anything? If not, it isn’t finished.
Getting the input right
Whatever the rig applies, it has to reach the device the way it would in use. For a force test that’s the load path, the route from the actuator through the fixture and the device into the base. A sample held slightly tilted sees a bending component the requirement never mentioned, and the numbers come out scattered or wrong. The same thinking applies to other inputs. A fluid rig with the wrong tubing bore or a trapped air pocket is loading the device with something real use wouldn’t.
A good share of design time therefore goes into features that make the set-up right by construction: dowels and shoulders so the sample sits square, fixed tubing runs, hard stops on stroke. Where a standard specifies orientation (the free-fall test in ISO 11608-1 drops each sample in set orientations, for instance), we design that into the rig so it’s fixed and documented, never set by eye.
Choosing the measurement
Whatever the rig measures (force, pressure, flow, displacement, temperature), the sensor has to be ranged so the expected readings sit comfortably within it. Most instruments are specified as a percentage of full scale, so a reading at 0.2% of range carries far more uncertainty than the requirement can tolerate. Measuring a 5 N cap removal force on a 2,500 N load cell gives you a number, in the sense that the software prints one, and not much else. Our Mecmesin MultiTest 2.5-dV takes interchangeable load cells, so for force work we pick per test, and the same logic applies to the pressure transducer in a fluid rig or the encoder on a cycling rig.
Calibration matters as much. Every measuring channel needs a current calibration certificate traceable to national standards, because a verification result measured on uncalibrated kit is easy for a reviewer to dismiss, and rightly so.
Single run or cycling
A single pull to failure answers “how strong is it”. Cycling answers “does it survive its service life”, which is usually the question the risk analysis raises. The rig implications differ. A single test can tolerate a rig that’s fiddly to load, because you load it once per sample. A test running ten thousand actuations usually justifies a dedicated rig with its own drive, a drive profile that matches real use (speed, stroke, dwell), hardware that survives the full count itself, and a way of detecting failure mid-test. We often add a witness measurement, such as a force reading every N cycles, so degradation shows up as it develops.
What we log
The number on the display is the smallest part of the record. For each test we log the full trace (force against displacement, pressure over time, whatever the rig produces), because the shape of the curve shows how the device behaved or failed. Alongside the traces we record sample identity and build standard, the rig’s identity and configuration (including any surrogates and the drawing issue they were made to), sensor identities with calibration dates, test settings, ambient conditions where they matter, and any anomalies, verbatim, however minor they seemed at the time.
A design verification result has to be reconstructable years later by someone who wasn’t in the room. If the record supports that, the test was worth running.
Drawings and a written method
The rig gets drawings and the test gets a written method, and those two documents are what separate a verification test from an informal bench check. The drawings mean the rig can be remade identically if a repeat is ever needed. The method (sample preparation, loading procedure, settings, acceptance criteria, data to be recorded) means someone else can repeat the test and get the same answer. Both go into the file alongside the results, and reviewers do read them.
We take on test rig design as part of our proof of concept and verification builds, usually alongside the prototypes the rigs are testing, which keeps the loop between a failed test and a revised part short.
FAQ
A test rig is purpose-built hardware that presents a device, or part of one, under defined conditions and measures its response, so a design requirement can be verified with repeatable, documented evidence. That can be a custom fixture in a universal force tester, a rig with its own actuation and sensing, a fluid circuit, or a surrogate for the tissue or product the device interacts with in use.
Every measuring element (load cells, pressure transducers, flow meters, displacement measurement) needs current calibration traceable to national standards. The rig as a whole should be checked before use to confirm it applies the intended input in the intended way, and that check should be recorded. Custom fixtures and surrogates don’t carry calibration certificates, but their drawings and identity should be documented in the test record.
Yes, provided the fixture is stiff enough that its own deflection doesn’t affect the measurement, and it survives the test without wear changing the sample fit. We use printed parts routinely for lower-load tests and early iterations, and move to metal for high loads or long cyclic tests.
It depends on the requirement and the statistical confidence you need. Standards such as ISO 11608-1 set out sample sizes for specific tests, and where no standard applies the sample size should be justified in the test method, typically from the risk associated with the requirement.
The hardware can be identical. Design verification is run against an approved method with defined acceptance criteria, on documented samples, with calibrated equipment, and the results go in the technical file. An informal bench test is for learning and is still worth doing early, but it can’t stand as verification evidence.
If you have a test requirement and no rig, or results you’re not confident would survive review, I’d be happy to have a call and talk through what the test rig design needs to cover. You can book a free 30 minute call here.